 | Electricity
Rated power consumption: 3 kW
Mains supply connection:
400/230 V; 50/60 Hz; 3P + N + E
Cold water
Tap water consumption: 225 l/h
Tap water quality: hardness between 8 and 30 French degrees (ideal 18 French degrees)
Tap water temperature: 10°C - 50°C
Tap water connection: A manually enabled valve and tee with ¾" G internal thread
De-ionized water
De-ionized water consumption:
About 20 l/h (closed circuit)
De-ionized water quality:
< 0.2 microSiemens x c
Water drain
Rinsing water drain / chemical tank drain: 1" (25 mm) tube
Others
Exhaust fan tubing: 4" (100 mm) (I.D.)
Main dimensions approx.
(wxdxh)
1500/800/1300 mm
59/32/51 inches
Weight
340 kg/750 lb |
|

|
|
 | |  |

 | Automated lift out of lenses |

|
|